Fluke® FLUKE-789 FLUKE-789 FC | Fromm
Send Print
  1. pin-int

Fluke® FLUKE-789

Fluke® FLUKE-789 Process Meter, 0 to 1000 VAC/VDC, 0 to 1 A, Multi-Line LCD Display, Plastic

  • Manufacturer Name: Fluke
  • MPN: FLUKE-789
  • Brand Name: Fluke®
  • UPC: 095969112192
  • FEPN: FLUKE-789 FC
  • Min Order Qty: 1
  • Quantity Interval: 1
  • Inventory Availability loading

The Fluke 787 was the first tool to combine a DMM and a loop calibrator to give process technicians double the power in one tool. Now it's even better, the range is extended with the Fluke 789 the ultimate loop calibration multimeter. The Fluke 789 has a 24-Volt loop supply reducing the need for taking a separate power supply when doing offline transmitter testing. And with its built-in, selectable 250-Ohm Hart® resistor, it also eliminates the need to carry a separate resistor with you. Now process technicians can do a lot more while carrying a lot less.

Fluke® Process Meter, 0 to 1000 VAC/VDC Loop, 0 to 1 A Loop, Multi-Line LCD Display, 7.35 in L x 3.41 in W x 1.25 in H Size, Plastic

Size       :      7.35 in L x 3.41 in W x 1.25 in H
Loop Voltage Rating       :      0 to 1000 VAC/VDC
Loop Amperage Rating       :      0 to 1 A
Display Type       :      Multi-Line LCD
Material       :      Plastic
  • 24 V loop power supply
  • Very large dual display
  • Enhanced backlight with 2 brightness settings
  • 1200 ohm drive capability on mA source
  • HART mode setting with loop power and a built-in 250 ohm resistor
  • 0% and 100 % buttons to toggle between 4 and 20 mA sourcing for a quick span check
  • Externally accessible fuses for easy replacement
  • Infrared-port to process data in FlukeView Forms (V2.1) software
  • Precision 1000 V, 440 mA true RMS digital Multimeter
  • 0.1% dc voltage accuracy, 0.05% dc current accuracy
  • Frequency measurement to 20 kHz
  • Min/max/average/hold/relative modes
  • Diode test and continuity beeper
  • Simultaneous mA and % of scale readout
  • 20 mA dc current source/loop calibrator/simulator
  • Manual step (100%, 25%, coarse, fine) plus auto step and auto ramp
  • Externally accessible battery for easy battery changes
  • Display current and % of span, auto step, auto ramp
  • -20 to 55 deg C operating, -40 to 60 deg C storage temperature, 95% upto 30 deg C, 75% upto 40 deg C, 45% upto 50 deg C, 35% upto 55 deg C humidity
  • pdfSpecification Sheet https://externalassets.unilogcorp.com/ASSETS/DOCUMENTS/ITEMS/EN/IDW_DOCUMENT_39fd49706b9d721947a6c5f6046745f3d23d6fb3ea1f6ddfdb47115d37e7fb60.pdf